Select TIME CONSTANT.
Press numeric key 2 and then the ENT key.
(3) The following display will be presented.
<Time constant setting screen>
TIME CONSTANT (0.01s=1, 0.02s=2, 0.05s=3,
4
0.1s=4, 0.5s=5, 1.0s=6, 2.0s=7)
Set a time constant value.
Press numeric key (1 to 7, usually input 4) and then the
ENT key.
* Default value: 4
(4) Constant setting is now completed and the 1
st
monitor
screen returns.
3.2.3 Use time program setting
It is selectable whether or not to use automatic change of
measurement parameters in response to measurement time
(time program).
(1) Press
the
UTLTY
SET UP
key.
(2) The following display will be presented.
<Utility initial screen>
UTILITY (OFFSET=1, TIME CONSTANT=2,
_3
USE PROG=3, PMT VOLT=4, OTHERS=5)
Select USE PROG.
Press numeric key 3 and then the ENT key.
(3) The following display will be presented.
<Time program setting screen>
USE TIME PROGRAM (YES=1, NO=0)
1
3 - 10
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...