2.5 Specifications
(7) Noise
(a) 1-wavelegth
measurement
:
0.5
×
10
-5
AU or less
(conditions: wavelength 250 and
600 nm, air in cell, time constant
2.0 sec, data sampling period 400
ms)
(b) 2-wavelegth
measurement
:
3.5
×
10
-5
AU or less
(conditions: wavelength 210 and
250 nm, air in cell, data sampling
period 800 ms)
(8) Drift
: 1.0
×
10
-4
AU/h or less
(conditions: wavelength 250 and 600 nm, air
in cell, constant temperature, time constant
2.0 sec)
(9) Auto zero : Zero adjusting range -0.2 to 2 AU
(10) Offset
: 0 to 2 AU, settable in 0.001 steps
(11) Flow cell : Quartz, SUS, Fluororesin (optical path length
10 mm, cell capacity 13
μ
L)
(12) Withstand pressure of flow cell
: 1.0 MPa
(13) Controllable flow cell temperature
(option)
:
40
°
C (environmental
temperature range: 4
to 30
°
C)
(14) Time program
No. of programs
: 9
Settable time
: Up to 600 minutes in increments of
0.1 min
No. of steps stored : 100 steps in total of 9 files
Program contents
: Measuring wavelengths, baseline
processing (usable only in 1-
wavelength measurement)
2 - 8
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...