4 - 20
(4) Cleaning
Microsyringe
(a) Wipe the needle before and after sampling.
(b) Wipe the needle before aspirating a cleaning solvent.
Clean the inside of the syringe seven times or more.
(c) Clean the syringe after injection immediately.
Fig. 4-8 Cleaning Microsyringe
(5) Tightening Connection Tube
If a Teflon tube is tightened too much, the inside diameter
becomes short, and the pressure increases abnormally.
When tightening a Teflon tube or a SUS pipe, see the
following.
(a) Push screw (resin screw); Tighten it fully with your
fingers (using no tools.)
(b) Push screw (SUS screw); Tighten it fully with your
fingers (using no tools), then re-tighten with a wrench
(with a tightening torque of 4 [N•m]) to secure firmly.
Refer to the table shown below about an angle of re-
tightening,
Table 4-4
Angle of Re-tightening
SUS pipe installation
To connect a new pipe
(Ferrule not yet caulked)
To reconnect an existent pipe
(Ferrule already caulked)
Angle of re-tightening (°)
270 to 360
30 to 60
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...