(b) Grounding Connection when grounded 3P power outlet
not available
2P power outlet
(15A)
Grounded 3P
table tap
Power cord
(standard
accessory)
Chromaster 5440
fluorescence detector
Ground
terminal
750 mm or
more
underground
Grounding
resistance:
100
Ω
or less
Connector
Ground electrode
(Underground)
Ground wire (green)
Screw diameter :
M4 or larger
Turn count
:
3 or more
Wire thickness
:
1.25 mm
(b-1) With 2P Power Outlet Connection
Rear panel
Copper plate
300 mm
2 mm
10 mm
Ground terminal
Power outlet having
a ground terminal
3P-to-2P adapter
Chromaster 5440
fluorescence detector
Connector
Rear panel
Ground wire (green)
Screw diameter :
M4 or larger
Turn count
:
3 or more
Wire thickness
:
1.25 mm
(b-2) With Power Outlet with Ground Terminal
Appended Fig. 1-2 Connection when Only 2P Power Outlet
is Available
APPENDIX 1 - 4
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...