All Volume Injection Method Procedure
The all volume injection method, the injection volume, the
feed volume, and the air volume are restricted by the following
relational expression. A stroke of the syringe capacity can’t
be used all.
<175 µL of Standard syringe >
vi + vf + va
≦
165 µL
vi :Injection volume
(The sample net volume to the flow path)
vf :Feed volume
(The wash fluid volume to deliver a sample to the loop)
va :Air volume
(The air volume of the lead and rear part of a sample)
100 µL loop is used for the autosampler as a standard. In
general, the feed volume is set as from 30
μ
l to 40 µL when
using 10
μ
l of sample.
In All volume injection method, the condition as [va: 0
μ
l of Air
volume] is recommended.
For a high accuracy gradient analysis, wash flow paths of the
buffer tube and the needle with the wash fluid2 after an
injection to avoid inflow of the wash fluid1 into the column.
2 - 26
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...