3.5.3 Liquid crystal display contrast setting
Set a contrast (brightness) level of the LCD screen.
1)
Turn on the power switch with the
UTLTY
SET UP
key held down.
2)
The following display will be presented.
<Setup initial screen>
SETUP: DETECT THERMO CONTRAST S. NO.
CH1
USE
4 02AB-111
Press the ENT key.
(3) The following display will be presented.
<Setup items selecting screen>
SETUP ITEMS (DETECTOR=1, CONTRAST=2
3
CONTRAST=3, EXCHANGE STANDARD=3)
Select CONTRAST.
Press numeric key 2 and then the ENT key.
(4) The following display will be presented.
<LCD contrast setting screen>
LCD CONTRAST (1 - 7)
3
Set a LCD CONTRAST value.
Press a numeric key (1 to 7, usually input 3) and then the
ENT key.
* 1:Dark
→
7:Bright
* Default value: 4
(5) LCD contrast setting is now completed and the setup initial
screen appears.
Under CONTRAST, the selected contrast value is indicated.
<Setup initial screen>
SETUP: DETECT THERMO CONTRAST S. NO.
CH1
USE
3 02AB-111
3 - 49
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...