(a) Sequential
parameter
The sequential parameter is settable in the following
procedure.
1 ENT
1 ENT
Input the start step (for start from a halfway step in a
sequential program).
NOTE
: The set value of start step will return to 1 once the
autosampler has been started by pressing the
START
key.
2 ENT
At the end of a sequential program, the ALL END signal can
be output. This signal is usable for control of other module
connected.
For output of the ALL END signal, input “1” (YES) or “0”
(NO) for avoiding the output.
0 ENT
UTILITY (SEQ=1, RACK=2, WASH PARAM=3,
1 OTHERS=4)
SEQ PARAM (START STEP=1, ALL END=2,
1
OPERATION=3, INJ TIMING=4)
START STEP (1-100)
1
Monitor screen reappears.
ALL END OUT (YES=1, NO=0)
1
Monitor screen reappears.
(to be continued)
3 - 18
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...