3.6.1 Selection of Mobile Phase and Cautions on Handling
(1) Be careful about restriction on use due to characteristics
and temperature of mobile phase to be used. (See Table
3-10 Characteristics of Organic Solvents)
(2) When performing high sensitivity analysis with a UV
detector, use solvents whose UV absorption characteristics
are especially small. Use solvents for liquid
chromatograph commercially available.
Absorption characteristics of special grade reagents such
as methanol, ethanol and acetonitrile generally and widely
used are greatly different in the short wavelength range,
depending on the manufacturing lot and the manufacturer.
Be careful especially when performing high sensitivity
analysis.
(3) Degas the mobile phase before use without fail.
(a) It is recommended to use a degassing unit because
organic solvents have a specially high content of air,
and the baselines may be fluctuated by bubbles.
Malfunction of check valve and loss of column
efficiency may also be caused. Degassing unit is
especially required for mixing solvents for gradient use.
Degas the solvents before use seeing 3.7.2 Degassing
Mobile Phase.
(b) When performing analysis with a UV detector in the
short wavelength range or highly sensitive analysis
with an RI detector, degassing the mobile phase
before use is not enough, and the baselines may be
fluctuated by the bubbles. In such a case, use
degassing unit which can degas mobile phase on-line.
(4) Mobile
Phase
When using mobile phase for liquid chromatograph, since
the connections of pipings of pump and separation column
use stainless steel (SUS316), solvents or reagents that
may corrode stainless steel may be restricted.
Especially, solvents or solutions which contain halogen
ions such as HCl, KCl, NaCl, NH
4
Cl are typical reagents.
Reagents which may be restricted are shown in Table 3-9
use the table as a rough standard.
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...