(continued)
3 ENT
APL PARAM (SYR SPD=1, NEEDLE SPD=2,
1 SYR VOL=3, AIR VOL=4)
Select an operation condition.
1 ENT
Set a syringe speed to aspirate a sample
according to the sample viscosity.
Set a slow syringe speed for a high-viscosity
sample.
4 ENT
DSP SYRINGE SPEED (1-5)
4
Set a syringe speed to discharge a sample
according to the sample viscosity.
Set slow syringe speed for a high-viscosity
sample.
4 ENT
Relation between Syringe Capacity and Syringe Speed
Syringe
Capacity
Speed
1
Speed
2
Speed
3
Speed
4
Speed
5
70 µL
0.4 µL/s
1 µL/s
4 µL/s
10 µL/s
35 µL/s
175 µL
1 µL/s
2.5 µL/s
10 µL/s
25 µL/s 87.5 µL/s
700 µL
1 µL/s
4 µL/s
10 µL/s
20 µL/s 87.5 µL/s
Monitor screen reappears.
ASP SYRINGE SPEED (1-5)
4
(to be continued)
3 - 21
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...