(b) Connection when Grounded Power Outlet is
Unavailable
(b-1) With 2P power outlet
750 mm or
more below
ground
Copper plate
2P power outlet
(15 A)
Grounding
terminal
Ground
resistance
100
Ω
or less
Grounding
electrode
(underground
section)
Grounding construction
for 100
Ω
or less
3P table tap with
grounding wire
Power cord
(attached to
instrument)
Grounding wire (green)
Screw diameter :
M4 or larger
Turn count
:
3 or more
Wire thickness
:
1.25 mm
Chromaster 5310
column oven
Connector
Rear panel
10 mm
2 mm
300 mm
(b-2) With 2P power outlet having a grounding terminal
Power outlet with
grounding terminal
Grounding
terminal
2P-3P adapter
Grounding wire (green)
Screw diameter :
M4 or larger
Turn count
:
3 or more
Wire thickness
:
1.25 mm
Chromaster 5310
column oven
Connector
Rear panel
Power cord
(attached to
instrument)
Appended Fig. 1-2 Connection when Only 2P Power Outlet
is Available
APPENDIX 1 - 6
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...