3 - 55
(5) Deletion of Step
How to delete one step in a program is shown next.
PROG
(SET PROG No.)
No. of program containing step to be
deleted (m)
ENT
(MODIFY=0,
NEW=1)
0
ENT
(MAX)
Upper limit pressure
level
ENT
(MIN)
Lower limit pressure
level
ENT
DEL
ENT
ESC
1
Program number can be from 1 to 9.
2
For editing, input 0. For generating a new program, input
1.
3
If pump pressure rises beyond the level set here, the pump
automatically stops.
Set pressure is from 0.1 to 60.0 MPa.
4
If pump pressure falls below the level set here, the pump
automatically stops.
Set pressure is from 0.0 to 60.0 MPa.
5
Press the
key until the step you want to delete appears
on the screen. If display advances beyond that step, press
the
key.
6
Press the DEL key when the cursor is on the TIME line
want to delete.
7 When the ESC key is pressed, Edit is finished, and
Monitor screen appears.
1
2
3
4
5
6
7
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...