(4) The following display will be presented.
<Logbook selecting screen>
LOGBOOK (REPORT=1, RESET=2)
1
Select Report.
Press numeric key 1 and then the ENT key.
(5) On the logbook report screen, the total on period of the Xe
lamp and its standard lifetime (in parentheses under Xe
LAMP), the number of switching-on times and the date of
the last lamp change will be indicated.
<Logbook report screen>
Xe LAMP
SWITCHING
LAST CHANGED DATE
123h (500)
23 times
9/ 1/2010
Press the ENT or ESC key.
(6) Logbook display is now completed and the 1
st
monitor
screen returns.
3.3.2 Xe lamp logbook data resetting
Total on period of the Xe lamp, the number of switching-on times
and the date of the last lamp change are reset.
(1) Press
the
GLP
key.
(2) The following display will be presented.
<GLP initial screen>
GLP (LOGBOOK=1, KEY LOCK=2,
1
WL CHECK=3, Xe CHECK=4)
Select LOGBOOK.
Press numeric key 1 and then the ENT key.
3 - 18
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...