Connect the syringe to the tip of the drain tube, and inject
the wash fluid into the cell for cleaning (Fig. 5-7).
NOTE
: This flow cell unit (P/N 890-2967) uses a part similar to
the one used in the Chromaster 5440 FL Detector.
Be careful not to mount the part of the flow cell unit to
the detector and vice versa.
Fig. 4-9 Flow Cell Unit
(6) Replace the wash fluid with the mobile phase that is used
for separation, and repeat the procedure in step (5).
(7) If air bubbles are observed in the flow cell, supply solution
into the cell with the pump at a flow rate within 1.0 mL/min
and a pressure within 1.0 MPa. While the solution is being
supplied into the flow cell, pinch the end of the drain tube
with your fingers for a few seconds to increase the pressure
inside the cell, then release your fingers quickly.
Repeat this procedure a few more times until all the air
bubbles have been removed.
4 - 25
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...