5 - 13
(4) Prepare the wash fluid in a beaker, and place the filter in it.
Place the beaker in the ultrasonic bath. Run the ultrasonic
cleaner for 5 to 6 minutes.
(5) Replace the wash fluid with distilled water. Run the
ultrasonic cleaner for an additional 5 to 6 minutes.
Repeat this operation three or more times to eliminate the
wash fluid.
(6) After removing the moisture from the filter, mount on the
inlet tube again.
Solvent filter
Beaker
Wash fluid
↓
Distilled water
Ultrasonic cleaner
Inlet tube
Insert
Solvent filter