(iii) Press the WL key and the wavelength setting
screen appears.
Set an excitation (Ex) wavelength.
<Wavelength setting screen (Ex indicated)>
Ex
Em
(200 - 850.0)
550 350
Press numeric keys 5 5 0 and then the ENT key.
(iv) Emission (Em) wavelength will not be set here.
Leave the present setting as it is.
<Wavelength setting screen (Em indicated)>
Ex
Em
(250 - 900.0)
550
350
Press the ENT key.
(v) The
1
st
monitor screen appears.
<1
st
monitor screen>
TIME Ex Em FL PROG
LAMP [I]
0.0
550 350 ****
1
Xe
(vi) In this status, place a paper strip at the Ex beam
exit center of the sample compartment as shown
in Fig. 5-9 and observe the condition of the
excitation beam.
Repeat the VERTICAL, HORIZONTAL and
FOCUS adjustments in this order little by little
until the excitation beam (rainbow color) becomes
brightest. The coarse adjustment of the light
source position is thus completed. Please adjust
VERTICAL first in installation because the lamp
position is shifted downward for fixation of the
transport screw.
5 - 15
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...