3.1.2
NOTE
: The time for the instrument to stabilize varies with
the measurement conditions. Secure a warm-up
time depending on the conditions.
(Reference warm-up time: About 1 hour)
NOTE
: Temperature of the thermostatic flow cell varies
with the set time (measuring conditions). If pump
flow velocity or oven temperature is changed,
provide a warmup time in accordance with
measuring conditions (reference: about 30
minutes).
3.1.2 Parameter Input Method
(1) Display Contents of Screen
Explained here are display contents of the parameter input
screen and program setup screen.
(a) Input of a single numeral
PROGRAM NO. (1-9)
1
Parameter name
Input range
Input part
(b) Input of plural numerals
LAMP CHANGE (MM, DD, YYYY)
<1-12> _
Parameter name
Input item
Input range
Input part
(c) Selection
of
numeral
TIME CONSTANT (0.01 s=1, 0.02 s=2, 0.05 s=3, 0.1 s=4
6
0.5 s=5, 1.0 s=6, 2.0 s=7)
Parameter name
Selectable values
Input part
3 - 5
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...