(3) Press
the
WASH
key of the autosampler, and select
NEEDLE
side to fill the needle and the injection port with
wash fluid1. (When two wash fluids are used, replace the
wash fluid2 side with fresh one). Fill the buffer tube with
deaerated wash fluid while flushing the needle and the
injection tube.
(4) Place a vial containing at least 1 mL of standard sample in
rack position No. 1 and set the rack in the autosampler.
(5) Press
the
START
key of autosampler to start analysis.
The analysis is carried out 6 times. The operating
conditions during analysis are indicated on the display.
Ensure beforehand that the room temperature will not
change suddenly during analysis.
(6) Upon completion of analysis, the operation of the system
stops with the initial screen displayed. Instead of the initial
screen, the ALL END screen will appear if ALL END is set.
In this case, the initial screen appears when you press the
IDLE
key.
(7) Data
Check
Determine the Relative Standard Deviation (RSD)
of peak
area by the following calculations.
Mean value
n
xi
x
n
1
i
∑
=
=
Standard deviation
1
n
)
x
xi
(
SD
n
1
i
2
−
−
=
∑
=
Relative Standard Deviation
(%)
100
×
=
x
SD
RSD
xi: Area value due to each injection,
n: Number of injections
Make sure the CV value is 0.2% or less (injection volume is
10 µL or less). However, this value represents the
reproducibility of not the autosampler alone but the entire
system. Therefore, the value varies with each stability of
the pump and detector, accuracy of the data processing
unit and its processing parameters.
4 - 6
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...