(6) S/N Ratio Calculation for Raman Spectrum of Water
Perform calculation according to the S and N obtained in
the preceding subsections (3) and (5).
(a) Excitation spectrum bandwidth: 15 nm
S/N ratio = S/N (baseline method, S/N ratio
≥
700)
S/N ratio = S/N (tangent method, S/N ratio
≥
525)
(b) Emission spectrum bandwidth: 30 nm
S/N ratio = S/N (baseline method, S/N ratio
≥
900)
Noise value =
Σ
Y
R/n
(Xn shall be determined for
intervals of 0.5 minute)
R=n
R=1
Fig. 4-3 Noise Calculation (extracted from
JAIMAS 0005-1984)
NOTE
: The Raman spectrum of water and the noise level
may be affected by the purity of the water used
and/or the presence of air bubbles.
(7) Measurement of Raman Spectrum of Water Using a Time
Program (when Raman signal intensity obtained in the
preceding subsection (2) is low)
If the sensitivity (S/N ratio of Raman peak of water) is low,
recheck the following items.
(a) When noise is high
(i)
Check the distilled water.
* Distilled water should be used soon after its
purification. If left for a long time in a
container made of resin or the like, fluorescent
substance may mix in with the distilled water.
4 - 11
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...