(b) Temperature
calibration
If there is a difference between the indicated
temperature of the thermostat and the actual
temperature of the liquid in a sample vial, it can be
corrected by setting of temperature calibration.
In case the indicated temperature of the thermostat is
10.0
°
C and the actual temperature of the liquid in a
sample vial as measured with a calibrated
thermometer is 10.5
°
C, the difference correction value
(offset value) is +0.5. So, setting is required as
follows.
THERMO
2 ENT
Select on which side there is a difference, plus side or minus
side.
e.g.) 1 ENT
Set an offset value for correction.
e.g.) 0.5 ENT
ENT
or
ESC
STEP RACK VIAL VOL(µL) INJ TIME [I]
1 1 1 10.0 1/1 0.0
THERMO UNIT (SET TEMP=1, CALIBRATION=2,
2 TOLERANCE=3, SERIAL NO.=4)
The presently set value is indicated.
SELECT MODE (+OFFSET=1, -OFFSET=2)
1 OFFSET(
°
C)= 0.0
SET +OFFSET (0.0-5.0)
°
C
0.0
OFFSET (
°
C)
0.5
Monitor screen reappears.
3 - 54
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...