SAFETY - 12
SAFETY SUMMARY
NOTICE precaution
(Continued)
Accuracy and Precision of Measured Values
Carry out periodic inspection and check whether the system is operating normally.
If necessary, conduct measurement on a control sample.
Possible Carryover
When examining the results of measurements, take account of the possibility of occurrence
of a carryover. Otherwise, a correct judgment may not be formed. Since fluororesin and
quartz materials are used for the wetted parts of piping, particular attention should be paid
when analyzing samples that are likely to be adsorbed by these materials.
Before Turning on Xenon Lamp Again
After the xenon lamp is turned off, it remains very hot for a while. Under this condition, it is
not allowed to turn on the xenon lamp again for measurement.
Before turning on the xenon lamp again, wait for at least five minutes.
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...