(f) Shall be free from excessive dust.
Otherwise, performance degradation might be caused.
(g) Shall be free from a sudden change of power supply
voltage. Otherwise, noise might be caused.
(h) Power supply to motor-driven equipment (such as
stirrer and vibrator) without a noise suppressor shall
not be turned ON/OFF frequently on the same power
supply line as that of this instrument.
3. Items to be Prepared by Customer
3.1 Power Supply Arrangement
(1) Power supply voltage: 100 to 115 V AC/220 to 240 V AC
Fluctuation shall be ±10% of the rated voltage.
(2) Frequency: 50 or 60 Hz
Fluctuation shall be ±0.5 Hz of the rated frequency.
(3) Power Capacity: 450 W
Confirm the necessary power capacity taking actual system
configuration and OA equipment in the installation room into
account and prepare a power supply having an adequate
power capacity.
(4) Power
outlet:
Power supply shall be obtainable near the place for
installing the system inclusive of this detector.
This detector requires a single grounded power outlet (3-
pole type). One or more outlets are needed for the entire
system.
APPENDIX 1 - 3
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...