(b) Check of wavelength accuracy
Check the wavelength accuracy by using the bright line
at 254 nm of the Hg lamp (see section 4.4.4 (6)).
“GOOD” will be indicated if the wavelength error is
within
±
3 nm, or “FAIL” will be indicated if the error
exceeds
±
3 nm.
NOTE
: First adjust the xenon lamp position (see 5.2.1
(3)) and then check the wavelength accuracy.
An accurate wavelength check cannot be
made unless the lamp position is adjusted.
NOTE
: If “FAIL” is indicated, wavelength calibration is
required. Refer to paragraph (2) below for
the calibration.
(2) Calibration of Wavelength Accuracy
If there is a wavelength deviation, then adjust the
wavelength in the wavelength calibration mode (special
mode).
(a) Calibration of excitation wavelength
(i) Press
the
UTLTY
SET UP
key, and the utility setting
screen appears.
<Utility setting screen>
UTILITY (OFFSET=1, TIME CONSTANT=2,
99
USE PROG=3, PMT VOLT=4, OTHERS=5)
Press numeric keys 9 9 and then the ENT key.
(ii) The excitation wavelength calibration mode
screen appears.
<Excitation wavelength calibration mode screen>
Ex WL ADJUST (1 - 20)
10 (ONLY FOR FACTORY USE) (-2)
The present offset value is indicated at Ex WL
ADJUST. Indicated at the bottom right of the
screen is the wavelength check result (deviation
of wavelength; indication range of –10 to 10).
4 - 16
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...