(x) The
1
st
monitor screen returns.
(d) Setting of spectrum measurement:
(i)
Press the SPECT key. The scan mode setting
screen appears.
<Scan mode setting screen>
SPECTRUM MODE (FL (Ex)=1, FL (Em)=2)
2
Press numeric key 2 and then the ENT key.
(ii) Set a scanning range.
➀
Enter a starting wavelength (minimum value)
for WL1.
<Scan range setting screen>
SCANNING WL RANGE (WL1 WL2) nm FL (Em)
<250 – 850>
350 - 700
Press numeric keys 3 5 0 and the ENT key.
➁
Enter an end wavelength (maximum value) for
WL2.
<Scan range setting screen>
SCANNING WL RANGE (WL1 WL2) nm FL (Em)
<
400 - 750
>
350
-
450
Press numeric keys 4 5 0 and the ENT key.
(iii) The memorize spectrum No. setting screen
appears.
<Memorize spectrum No. setting screen>
MEMORIZE SPECTRUM NO. (1 - 4)
1
Press a numeric key (1 to 4) and the ENT key.
(iv) The scan monitor screen appears, and spectrum
acquisition begins.
4 - 6
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...