(c) HEIGHT
The needle can be moved to the specified height.
STEP RACK VIAL VOL(µL) INJ TIME [I]
1 1 1 10.0 1/1 0.0
FUNC
FUNC (TUBE=1, MOVE=2, HEIGHT=3, SYR=4,
3 VALVE=5, SVALVE=6, SPURGE=7, WPUMP=8)
3 ENT
ESC
HEIGHT ( Z SPEED)
<0.0-53.0> 0.0 2
HEIGHT ( Z SPEED)
<1-2> 14.5 2
e.g.) 14.5 ENT
MOVING [B]
WAIT!
2 ENT
IDLE
PRESS FUNC KEY OR IDLE KEY [W]
FUNC
ESC
ESC
Input a desired height.
Input a needle moving speed.
1: The needle will move at the low speed. Select this speed
for using a hard septum, etc.
2: The needle will move at the high speed. Usually, this
speed is used. However, for moving the needle upward,
the needle will move at the high speed irrespective of the
setting made here.
The needle moves to the specified height.
MOVING [B]
IDLE WAIT!
Monitor screen reappears.
3 - 44
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...