3.3 GLP-Related Functions
3.3.4 Checking the Energy of W Lamp
Energy of the W lamp is checked using the 600 nm bright line of
W lamp.
1)
Press key.
GLP
2)
The GLP setup screen appears.
GLP Setup Screen
GLP (LOGBOOK=1, KEY LOCK=2,
1
D2 CHECK=3, W CHECK=4, Hg CHECK=5)
Select W
CHECK.
Press [numerical] key [4], and then the [ENT] key.
3)
Checking action will automatically be taken.
During check, the following display is presented.
CHECKING LAMP ENERGY
(W)
4)
After this check, its result is indicated.
W
ENERGY
405775
The indicated energy value corresponds to the reference
side at 600 nm.
3 - 28