2.1
2. FUNCTIONS
2.1 Front Panel
6-channel degassing unit
Monitor lamp
Flow path inlet of degassing chamber
Flow path outlet of degassing chamber
Fig. 2-1 Front View of Pump
(1) Monitor
Lamp
Listed below are status indications of the monitor lamp.
The lit-state color of this lamp changes in response to the
status of vacuum pump.
Lamp Color
Vacuum Pump Status
Orange
(blinking)
Vacuum pump begins running upon turning on
power supply.
Green
(blinking)
Vacuum level has reached the upper limit of
control range.
Green
Vacuum level has reached the set control
point.
Orange
The vacuum pump has stopped due to
occurrence of an error.
(2) Flow Path Inlet/Outlet
Inlets and outlets for solvents before and after degassing.
Because flow paths are independent of one another, the
inlet and outlet of the same flow path must be paired.
2 - 1
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...