3.4.1 Selection of Mobile Phase and Cautions on Handling
(1) Be careful about restriction on use due to characteristics
and temperature of mobile phase to be used. (See Table
3-5 Characteristics of Organic Solvents)
(2) Mobile
phase
(a) When performing high sensitivity analysis with a UV
detector, use solvents whose UV absorption
characteristics are especially small. Use solvents for
liquid chromatograph commercially available.
Absorption characteristics of special grade reagents
such as methanol, ethanol and acetonitrile generally
and widely used are greatly different in the short
wavelength range, depending on the manufacturing lot
and the manufacturer. Be careful especially when
performing high sensitivity analysis.
(b) When using mobile phase for liquid chromatograph,
since the connections of pipings of pump and
separation column use stainless steel (SUS316),
solvents or reagents that may corrode stainless steel
may be restricted.
Especially, solvents or solutions which contain halogen
ions such as HCl, KCl, NaCl, NH
4
Cl are typical
reagents. Reagents which may be restricted are
shown in Table 3-2 use the table as a rough standard.
(c) Degas the mobile phase before use without fail.
It is recommended to use a degassing unit because
organic solvents have a specially high content of air,
and the baselines may be fluctuated by bubbles.
Malfunction of check valve and loss of column
efficiency may also be caused. Degassing unit is
especially required for mixing solvents for gradient use.
Degas the solvents before use seeing 3.4.2 Degassing
Mobile Phase.
(3) Wash
fluid
(a) Before using a wash fluid, be sure to degas it. If the
degassing of the wash fluid is inadequate, air bubbles
will occur in the flow path of the autosampler, resulting
in a decrease in reproducibility (Refer to 3.4.2)
3 - 61
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...