4.4.4
(2) Disassemble the parts, referring to the flow cell
disassembling diagram shown below.
Clamp screw 1
(655-1899)
Window support 2
(893-4427)(M11)
Spacer 2 (890-2958)
Cell lid (893-4409)
Window support 1
(893-4427)(M11)
Spacer 1 (890-2958)
Cell window (885-2832)
Packing 7.8
×
2.4 (885-2836)
Pressure withstanding cell
(893-4404)
Sleeve (L369053)
Cell window (885-2832)
Packing 7.8
×
2.4
(885-2836)
Clamp screw 2
(655-1458)(1/4 inch)
Liquid leaking hole
Fastening screw (M3)
Sleeve (638-1099)
Fig. 4-5 Disassembly of Flow Cell Main Frame
(3) Wash all the parts of flow cell with an ultrasonic washer.
(4) Wipe off contaminants from the surface of the flow cell with
gauze soaked in acetone or ethanol.
(5) After washing, assemble the parts in reverse to the
disassembly, referring to the disassembling diagram.
The pressure withstanding cell has holes for liquid leaking.
The cell should be assembled so that the holes are
positioned at the bottom.
(6) After assembling, flow liquid through the cell to confirm that
liquid does not leak.
4 - 17
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...