(1) Press
the
PROG
key.
(2) The time program No. setting screen appears.
<Time program No. setting screen>
SET PROGRAM NO. (1-9)
1
Press a numeric key 1 to 9 and then the ENT key to create
or change the program No., or select the existing program
No.
* Initial (default) value: 1
* If the setting will not be changed, press the ESC key to
return to the 1st monitor screen.
(3) The mode setting screen appears. “0” is always indicated
here.
<Mode setting screen>
EDIT MODE (MODIFY=0, NEW=1)
1
Press a numeric key 0 or 1 and then the ENT key to either
use (MODIFY=0) or delete (NEW=1) the conditions in the
time program No. selected above.
* 0 : Conditions in the time program No. selected above
are used; additions or changes can be made in the
program.
1 : Conditions in the time program No. selected above
are deleted, and new conditions can be set.
The system now advances to the step setting screen for
setting the time program, and setting is allowed for the first
step (the cursor moves to TIME).
3 - 28
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...