(2) Connect the provided hypodermic syringe to the port1.
Hypodermic syringe
Fig. 5-18
Connecting the Hypodermic Syringe (or Tube)
(3) Using the WPUMP command, deliver wash fluid1 to replace
the fluid of the flow path between the solvent filter and the
wash port with the wash fluid1.
FUNC
8
ENTER
1
ENTER
(Refer to 3.2.2 Analyzing Operation (5) Functions (h)
WPUMP for details
NOTES
:
To suspend delivering of the wash pump, select
OFF in the WPUMP PURGE screen, or press
the IDLE key. Delivering fluid is not suspended
automatically.
(4) After the fluid delivering by the wash pump, press the
hypodermic syringe (plastic injector) with your finger to
apply proper pressure. Then air bubbles can be removed
with ease.
Hypodermic syringe
Fig. 5-19
Discharge of Air Bubbles from Wash Pump with
Hypodermic Syringe
5 - 22
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...