3.4.4
(a) Input of Time
Press [numerical] key (within [0] to [6][0][0]), and then the
[ENT] key.
* Enterable range: 0.0 to 600.0 minutes, enterable in
steps of 0.1 minute.
(The first step is fixed at 0.0 minute.)
Upon pressing the ENT key, the cursor moves to the
measuring wavelength.
(b) Input of Measuring Wavelength
Press [numerical] keys (within [1][9][0] to [9][0][0]), and then
the [ENT] key.
* Enterable range: 190 to 900 nm, enterable in steps of
1 nm.
Upon pressing the ENT key, the cursor moves to BASE.
(c) Baseline
Processing
Press [numerical] key ([1] or [2]), and then the [ENT] key.
Input Numeral
Function
1
A/Z
: Auto zero will be executed.
2
HOLD : Baseline will not deviate.
Pressing the ENT key displays the next step.
Function of other keys during time program editing:
1) (clear)
: Clears an input numeral.
2) (delete) : Deletes a step.
CL
DEL
3) (insert) : Inserts a new step.
INSERT
PROG
4)
: Display the previous and next
steps, respectively.
▽
△
5) (escape) : Returns display to the monitor
screen.
ESC
LOCK OFF
3 - 39
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...