3 - 18
3.2 UTLTY Setting
3.2.1 Zero Pressure Calibration
The zero pressure point may change a little due to a secular
change of the pressure sensor. Please check the zero
pressure point once a month (Stop the pump, and then open the
drain valve). Refer to the following operation for calibration.
Monitor screen
PRESS is displayed as 0, and the zero pressure calibration is
completed.
UTLITY (CONDITION=1, UNIT=2, FLUSH=3
1 EVENT=4, LINK=5, P.WASH=6, OTHERS=7)
CONDITION SET (Z-ADJ=1, MOTOR-CTRL=2
1 PURGE=3, CALIBRATION=4, LPGR=5)
Select CONDITION
1 ENT
PRESSURE ZERO ADJUT (YES=1, NO=0)
1 (NOTE: OPEN DRAIN VLAVE)
Select Z-ADJ (ZERO ADJUST)
1 ENT
Select YES
1 ENT
Monitor screen
SET UP
UTLTY
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...