Appendix 1.5.5 System Tubing
In the Chromaster series, the inlet tube up to the pump and drain
tube are fastened to the left side face of each unit with tube
clamps. And the tubes are connected from the pump to the
autosampler, column and detector through the central hole of
each module unit.
Drain tray
Hole on the right hand
Through this hole, the tube
from the autosampler to the
column is passed.
Central hole
Through this hole, the tube from the
pump to the autosampler and that from
the column to the detector are passed.
NOTE
: Please ensure that the detergent vial is located
higher than that the autosampler main unit. If it
is located lower than the autosampler main unit, it
is possible that the instrument is prevented from
achieving sufficient performance due to the
bubble entering the syringe because of the
negative pressure in the tubing.
NOTE
: Please ensure that the detergent is connected to
the degassing unit.
APPENDIX 1 - 20
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...