Retaining screws
Fig. 5-11 Removal of Light Source Cover
(c) Loosen the screw retaining the mercury lamp cover
(Fig. 5-12).
Hg lamp cover
Retaining screw
Fig. 5-12 Removal of Hg Lamp Cover
(d) Pull the Hg lamp cover straight out, and the Hg lamp
will be seen attached at the right side interior of the
main body (Fig. 5-13).
5 - 18