Transport
screw
Xenon lamp
Lamp house
Cathode nut
Cathode
nut
Appended Fig. 1-4 Transport Screws and Mounting Xenon
Lamp
(4) Install the xenon lamp in the lamp house. And mount it to
the instrument (see
⑧
to
⑪
in section 6.2.1 (2).)
(5) Adjust the position of the xenon lamp.
The xenon lamp position had drastically moved to the
downward of the vertical direction to fix the transport screws.
For this reason, adjust VERTICAL first (see section 6.2.1
(3).)
NOTE
: After removing the transport screws, please store
them.
APPENDIX 1 - 8