5.2 Lamp Replacement
5.2.1 Replacement of D
2
Lamp
(1) Part and Tools to be Prepared
•
New
D
2
lamp
•
Phillips
screwdriver
•
Blade-edged
screwdriver
(2) Check the following.
•
The power switch of this instrument is turned off.
•
The power cord is unplugged from the instrument.
(3) How to Replace D
2
Lamp
(a) Detach the lamp cover (by removing 2 fastening
screws).
Fastening screw
c
Lamp cover
(b) Loosen three screws of D
2
lamp lead wires connected
on the terminal block and pull out the wires.
Fastening screw
Terminal block
Lead wire
5 - 4