3.1.1
3. REPLACEMENT PARTS
3.1 Replacement Procedure
3.1.1 Fuse
Replacement
WARNING
Electric Shock due to Contact with Inside of Instrument
When replacing the fuse with a new one, be careful not to
receive an electric shock. To prevent this, be sure to turn
off the power switch before proceeding to fuse
replacement.
(1) Item to be Prepared
New fuse (Part No. J821337)
Be sure to use a time-lag fuse having the specified rating
(T6.3AL for 250 V).
(2) Pre-check
Condition
Make sure that the power switch of the instrument is turned
off and that the power cord is disconnected from the
instrument.
(3) How to Replace the Fuse
(a) Remove the fuse holder by hooking a tool having a thin
and flat tip (such as blade-edge screwdriver) to the
rear side of the fuse holder as shown below.
Fuse holder
Fig. 3-1 Removal of Fuse Holder
3 - 1
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...