6.3 Vials
NOTE
:
For performance specification checkup, use the
following parts in combination.
Part No.
Part Name
080-3595
1.5 mL sample vial
855-2351
Septum (Si -TEF), slit (-)
080-3704 Screw
cap
In the other cases, adjust the rack parameters (refer to
3.2) according to the combination used for your
practical application (indicated in Table 6-4 to 6-11)
and check the reproducibility performance before
proceeding to operation.
NOTE
:
For good reproducibility, avoid using a septum
repeatedly to prevent a cause of a hole. It causes a
fluctuation of sample concentration in sample vial.
NOTE
:
Be sure to use the sample vials, caps and septa
recommended by Hitachi High-Technologies (indicated
in Table 6-4 to 6-11).
If other parts are used, the instrument may malfunction
or its original performance may not be available.
6 - 3
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...