4.5 Error
Messages
4.5.1 Autosampler Main Unit Errors
Table 4-3 Autosampler Main Unit Errors
No.
Error Message
Possible Cause
Remedy
1 ROM
ERROR
2 RAM
ERROR
3 LOG_INFORMATION
ERROR
4 EEPROM
ERROR
5 PARAMETER
ERROR
PC board malfunctions.
Turn on power supply again.
If the same error is indicated
again, contact your nearest
service office of Hitachi High-
Technologies Corporation sales
representative.
6 COMMUNICATION
ERROR
Communication is abnormal.
Check connection of the
communication cable, and
reconnect the autosampler with
the chromatography data station
(CDS).
7
CHECK VIAL NO.
When starting analysis
(pressing the
START
key),
vial numbers beyond the
maximum number of vials as
a rack parameter are set in a
sequential program.
8 CHECK
START
STEP NO.
1. When starting analysis
(pressing the
START
key),
a start step No. which is
not contained in a
sequential program is set.
2. Setting of a start step No.
which is not contained in a
sequential program was
attempted by the UTILITY
function.
9
CHECK TOTAL VOL
An injection volume that
cannot be handled with the
mounted syringe is set.
Check and correct the set
parameter value.
10
START KEY IS NOT
ACCEPTABLEDOOR
IS OPEN
When starting analysis
(pressing the
START
key),
the front doors (R)(L) of the
instrument are unclosed.
Close the front doors (R)(L) of the
instrument.
11 SYSTEM_BUSY:
CHECK OTHER
MODULE STATUS
When starting analysis
(pressing the
START
key),
Other unit connected is in the
busy status.
Check status of other units and
release the busy status. (Stop of
liquid delivery with the pump,
setting of oven in condition before
heating ready status, etc.)
12 STOPPED
BY
LEAKAGE
1. Solvent is leaking through
the flow path of
instrument.
2. Solvent remains in the
drain.
1. Wipe off leaked fluid
completely and retighten the
pipe joint.
2. Wipe off remaining solvent
completely.
4 - 19
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...