NOTES
: 1. Be sure to measure the liquid temperature with a
calibrated thermometer before execution of a
temperature calibration.
2. A difference from the liquid temperature in a
sample vial varies slightly with the set temperature.
For temperature calibration, it is recommended to
determine the difference at the set temperature to
be used most frequently. And for temperature
calibration on multiple thermostats, be sure to
correct all thermostats at the same temperature.
3. In the thermostat, its aluminum block is subjected
to temperature control.
Therefore, a temperature difference occurs
between the set temperature and the liquid
temperature in a sample vial. If it is necessary to
control the absolute liquid temperature, you should
measure the liquid temperature in a sample vial
after reaching the set temperature and carry out
temperature calibration so that the target
temperature can be attained.
3 - 55
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...