3.1.3
3.1.3 Temperature
Setting
Carry out setting of the control temperature and temperature
upper limit.
SET TMP
Monitor screen
Set the temperature upper limit.
70 ENT
<45 - 95>
40
70
SET TEMP (
°
C)
MAX LIMIT (
°
C)
<1 - 85>
40 70
SET TEMP (
°
C)
MAX LIMIT (
°
C)
Set the control temperature.
40 ENT
NOTE
:
1. Unless a control temperature and temperature upper
limit are entered (and ENT key clicked) on the
temperature setting screen, the settings will not be
updated.
2. The temperature upper limiter should normally be set
at 70
°
C or higher except for a special reason.
If the limiter were erroneously set near the ambient
temperature, then a temperature limiter error may be
caused by ambient temperature fluctuation.
3. If a temperature upper limiter error occurs, the
temperature control will be stopped automatically.
To resume the temperature control, make resetting
on the temperature setting screen.
3 - 7
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...