5. PARTS REPLACEMENT
The following parts are consumables. Always prepare them for
replacement. Quantity of use is on a basis of rough standard,
assuming that the working days are 240 days a year, the working
hours are 8 hours a day. Normal life means usable time of the
part when the instrument is used with methanol flowing
(1.0 mL/min). However, since the values vary greatly depending
on the conditions (solvent, flow rate, pressure), prepare enough
quantity of each part, considering the actual conditions of use.
5.1 Pump Main Unit
Part No.
Name
Shape and Specifications
Position Used
Normal
Life
891-1213 Plunger
seal
(Pump seal)
Color: Black
Pump head for
Chromaster 5160
(4 months)
640 h
885-1884 Wash
mechanism seal
Wash mechanism
plunger guide
(4 month)
640 h
810-1163 Plunger
Pump head for
Chromaster 5160
1 - 3 year
885-2920 Packing
(Purge valve
seal)
Purge valve
(1 year)
1920 h
810-1128 Packing
Purge valve
(1 year)
1920 h
891-1228 Packing(PEEK)
Pressure sensor
holder
(1 year)
1920 h
Packing
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...