3.5 Setups
3.5.2 Setting of Thermostatic Flow Cell
Press and keys simultaneously.
UTLTY
SET UP
POWER
The SETUP screen now appears.
SETUP:
DETECTOR
THERMO
CONTRAST S.NO.
CH1 NO
USE
4
1401-001
Press the [ENT] key.
The selection screen appears.
SETUP ITEMS (DETECTOR=1, THERMO CELL=2,
1
CONTRAST=3)
Press numerical key [2], and then the [ENT] key.
THERMO CELL (NO USE=0, USE=1)
0
For avoiding use of
thermostatic flow cell : Press [numerical] key [0], and then the
[ENT] key.
For selecting use of
thermostatic flow cell : Press [numerical] key [1], and then the
[ENT] key.
After setting use or non-use of the thermostatic flow cell, “ON” or
“OFF” is indicated on the monitor screen 2.
WL SAMP
REF
CELL
[I]
250 301000 492000 ON
When the thermostatic flow cell is not yet connected,
“---“ will be indicated on the monitor screen 2.
WL SAMP
REF
CELL
[I]
250 301000 492000 ---
New setting is accepted after turning on power supply again.
NOTE
: Set the thermostatic flow cell in accordance with
measuring conditions. (Recommended flow rate
range when using thermostatic flow cell: 2mL/min
or less).
3 - 46
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...