5 - 11
(4) If a clogged filter has to be replaced, allow the filter has to
be replaced, allow the filter assembly to dry, then place it in
a freezer (-10
℃
) for about five minutes. The cold
temperature will contract the Teflon
®
packing and facilitate
removal of the filter.
(5) Carefully push and mount the new filter into the flow path
filter.
(6) Before attaching the flow path filter to the drain valve, clean
the flow path by running the pump.
(7) Attach the flow path filter to the lower right (canted) part of
the drain valve. Align the flow path filter with the drain valve
and tighten it securely with the wrench 21.
NOTE
: Poor tightening may cause leaks.
Flow path filter
Freezer (-10 °C about 5 minutes)
Filter
638-1423
Seal
638-1422
Flow path filter
Packing(PEEK)
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...