3 - 44
3.3.5 Low Gradient Unit Logbook
The total number of times of the changeover after replace of a
proportioning valve for a low gradient unit (option) is indicated.
This function is enabled only when the low gradient unit is
connected. This function is used as an index of a replacement
for maintenance parts of the low gradient unit.
(to be continued)
GLP (LOGBOOK=1, PRESSURE M=2, KEY LOCK=3)
1
LOGBOOK (PUMP SEAL=1, P.WASH=2,
5 A.VALVE=3, D.MIX=4, P.VALVE=5)
Select LOGBOOK
1 ENT
LOGBOOK (REPORT=1, RESET=2)
1
Select P.VALVE
5 ENT
GLP
Monitor screen
P.VALVE(A) LAST CHANGED DATE
10000000 09/01/2010
The total number of times of the changeover after
replacement of a maintenance part for the
proportioning valve is indicated.
ENT or ESC
1=Displaying logbook
2=Resetting logbook
Select REPORT
1 ENT
Monitor screen
REPORT (A=1, B=2, C=3, D=4)
1
Select a proportioning valve
1 ENT
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...