Raman spectrum of
water
Fig. 4-5 Drift Measurement
NOTE
: Carry out the drift measurement after the xenon
lamp has stabilized (at least one hour after the
lamp is turned on).
4.2.4 Wavelength accuracy check and calibration
(1) Check of Wavelength Accuracy
(a) Principle of measurement
Wavelength accuracy of the emission side
monochromator is checked by using the bright line at
254 nm of the builtin Hg lamp.
As for the wavelength accuracy of the excitation side
monochromator, the Em side is fixed at the wavelength
position where the abovementioned bright line was
detected, then the Ex wavelength where a Rayleigh
scattering peak of 254 nm is produced is measured.
D
Baseline (emission
wavelength 397 nm)
15 minutes
S (tangent
method)
Baseline
(baseline
method)
A
S
Drift:
D/A x 100
≤
3%/15min
Max.
wavelength
450 nm
Emission
wavelengths
Conditions:
Excitation(Ex)
wavelength ..................... 350 nm
Response ....................... 2 seconds
Flow cell.......................... 12
μ
L
Distilled water in cell
Room temperature kept constant
4 - 15
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...