4 - 7
4.2 Periodic Maintenance
4.2.1 Liquid Leak Sensor Check
The Chromaster 5160 pump is equipped with a liquid leak
sensor for detecting liquid leakage. Note, however, that liquid
leakage may not be detected correctly due to sensitivity of the
leak sensor. To prevent this, perform periodic operation check
for the liquid leak sensor as instructed below.
(1) Operation
check
Put 0.5 mL of water on the liquid leak sensor. And confirm
the leak error.
Fig. 4-2 Leak Sensor Position
NOTES
: 1. When a solvent is used, turn on the liquid leak
sensor function.
2. The liquid leak sensor may not detect a drop of
leakage or high-volatile solvents. Always check
the leakage even while the liquid leak sensor is
turned ON.
3. Visual confirmation of a liquid leak is
recommended as a daily check.
4. A periodic check for the liquid leak sensor is
required also.
Put 0.5 mL of water
Leak sensor
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...