Lamp igniter
+HV
COM
Cathode cable
Anode cable
Grip
Xenon lamp
Fig. 5-4
Lamp House after Removal
(f) Removal of old xenon lamp
WARNING
Injury due to Xenon Lamp Explosion (at disposal)
The xenon lamp is filled with high-pressure gas (approx.
1 MPa at room temperature, approx. 4 MPa under
operating condition), and this high-pressure gas still
remains in the xenon lamp after it is demounted for
replacement.
For disposal of the used xenon lamp, wrap it with a thick
cloth (e.g., triple-folded cotton cloth) completely and crush
its glass part with a hammer or the like. Then, discard the
xenon lamp as a dangerous waste item properly.
If the xenon lamp is discarded without being crushed, it
could explode due to possible impact at the time of
disposal, scattering glass pieces to cause personal injury.
WARNING
Injury due to Xenon Lamp Explosion (at mounting)
When loosening or tightening the retaining nut for the
xenon lamp, be careful not to apply excessive force to its
glass bulb part. Never hold and turn the glass bulb part
for loosening/tightening.
5 - 7
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...