2.3.2 Advantages of fluorometry
For verifying the advantage of fluorometry, the limitation of
absorbance measurement in its application to low-concentration
samples is explained first.
In this discussion, we will consider the detection of a sample that
has a transmittance of 99% (relative to the blank). If we
assume that the inaccuracy of the % transmittance measurement
is 0.1%, the reliability of the observed data is as follows:
Percent transmittance of blank
Percent transmittance of sample
100.0
±
0.1%
99.0
±
0.1%
Difference (proportional to concentration of sample)
1.0
±
0.2%
In this example, the uncertainty in the concentration
measurement is
±
20%. In contrast, in fluorometry, a difference
in the signal is directly proportional to the concentration of
sample.
An analysis of the error in the measurement is as follows:
Output signal level at measurement of sample
Value corresponding to blank
100
±
0.1
0
±
0.1
Difference (proportional to concentration of sample)
100
±
0.2
As is evident from this example, the error percentage by
fluorometry is not dependent on sample concentration
theoretically, so this method is very advantageous for low sample
concentrations.
In actuality, some error factors will increase relatively as sample
concentration falls extremely, but it can generally be said that
fluorometry is capable of detecting concentrations at least 3
digits lower than by absorptiometry.
2 - 8
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...