3 - 4
Table 3-3 2
nd
Monitor Screen
Indication
Description
P-UNIT
Pressure unit (bar, psi, kgf/cm2, MPa can be
selected). Use UTLTY to select a pressure
unit.
P.MAX
Maximum operating pressure.
When the pressure becomes higher than the
setup value, a pump is stopped
automatically.
P.MIN
Minimum operating pressure
When the pressure becomes lower than the
setup value, a pump is stopped
automatically.
D.MIX Dynamic
mixer
USE/No USE
P.WASH Wash
pump
ON: Running, OFF: Suspended
NOTE
:
For Chromaster 5160 Pump, an internal maximum
operating pressure will be set up according to the flow
rate setting.
Pressure limit range:
0 to 2.500 mL/min ............ 60.0 MPa,
2.501 to 5.000 mL/min ..... 30.0 MPa.
(d) 3
rd
Monitor Screen
Table 3-4 3
rd
Monitor Screen
Indication Description
LOW GR.
Low pressure gradient unit connection
state
P.WASH
Plunger wash pump connection state
A.VALVE
Auto purge valve connection state
D.MIX
Dynamic mixer connection state
OPTION LOW GR. P.WASH A.VALVE D.MIX [ -- ]
YES YES YES YES 1
Содержание CHROMASTER 5160
Страница 54: ...3 17 Fig 3 3b Connecting with the Chromaster 5260 Autosampler Injection valve Syringe valve Port4 ...
Страница 165: ...INDEX 3 U UI Pad 2 6 UTLTY 2 7 3 18 W Wiring covers 2 4 APPENDIX 1 19 Z Zero pressure calibration 3 18 ...
Страница 190: ...APPENDIX 3 2 Log book Means the information containing records of instrument operating conditions ...
Страница 221: ...3 1 Replacement Procedure b Replace the fuse with a new one and then attach the fuse holder by pressing it 3 2 ...
Страница 244: ......
Страница 406: ......
Страница 433: ...SAFETY 10 SAFETY SUMMARY WARNING and CAUTION Labels on Instrument 7 Before Turning on Xenon Lamp Front View ...
Страница 999: ...Retaining screws Fixing bracket Appended Fig 1 5 Removal of Fixing Brackets APPENDIX 1 13 ...
Страница 1016: ......
Страница 1022: ...Sample rack A component part on which a sample is set Logbook A historical record of instrument operation APPENDIX 3 2 ...